Copyright ©ERS Journals Ltd 2008 Validation of assays for inflammatory mediators in exhaled breath condensate1 Dept of Medicine, University of Birmingham, and 2 Queen Elizabeth Hospital, Birmingham, UK. CORRESPONDENCE: D. L. Bayley, Dept of Medicine, University of Birmingham, Edgbaston, Birmingham, B15 2TH, UK. Fax: 44 1216272012. E-mail: bayleydl{at}hotmail.com
Keywords:
Received: July 3, 2007
The use of exhaled breath condensate (EBC) as a tool for noninvasive assessment of lung inflammation is becoming commonplace. Many authors use commercial ELISA kits to measure inflammatory mediators within EBC. However, the very low concentrations of mediators within EBC are often below the commercially validated concentration range of the relevant ELISA and crucially below the linear part of the sigmoid standard curve. The present study seeks to validate a series of assays for use in EBC and to compare the results in EBC with those from matched sol phase sputum samples.
The following mediators were measured by ELISA: leukotriene (LT)B4, interleukin (IL)-8, secretory leukoprotease inhibitor and Mediator concentrations reached the lower limit of quantification in only one assay (AAT) in 19.6% of subjects, while mediator concentrations reached the lower limit of detection in three assays (LTB4, IL-8 and AAT in 31, 6.5 and 61% of subjects, respectively). No significant correlations were present between any mediators in EBC and sol phase sputum. The results of the present study indicate that care must be exercised when interpreting mediator measurements in exhaled breath condensate and that assays must be validated at concentrations relevant to those found within the biological fluid. Recently, there has been widespread interest in the use of exhaled breath condensate (EBC) as a fully noninvasive methodology for the assessment of inflammation in a variety of lung diseases, including chronic obstructive pulmonary disease (COPD), cystic fibrosis, bronchiectasis, primary ciliary dyskinesia and asthma. Analysis of EBC has used a variety of sensitive methods, including ELISA 1, multiplex ELISAs 1 and, more recently, mass spectroscopy 2. However, most investigators have used commercial ELISAs to determine mediator concentrations as they have the advantage of easy availability, simple methodologies and good reproducibility, and come "ready validated". Assay validation relates to a number of performance criteria, such as the lower limit of detection (LLD), reproducibility as determined by the coefficient of variation (CV; both intra- and interassay), linearity, spike return, performance in the media being assessed and specificity. However, commercial validation procedures do not describe the performance aspects of the assay within the relevant biological samples for differing disease states. Furthermore, the validation does not define lower limit of quantification (LLQ), which will also vary according to the biological fluid and the research requirements. The very low mediator concentrations commonly reported in EBC may, therefore, require a sensitivity not appreciated in the validation of a conventional commercial ELISA. This is especially important as the conventional validation of ELISA reproducibility is often only undertaken at concentrations on the linear section of the typical sigmoid curve for mediator concentration to signal. The variability of measurements increases greatly outside this linear portion and is particularly relevant for measurements at or around the LLD.
A variety of inflammatory mediators have previously been assessed in EBC in a range of lung diseases. Most interest has been focused upon the eicoinasoid leukotriene (LT)B4, typically measured using a commercial ELISA from Cayman Chemical Company (Ann Arbor, MI, USA). Concentrations reported in stable COPD range from a median value of 10.6 pg·mL–1 3 to 100.6 pg·mL–1 1. Izquierdo et al. 4 reported LTB4 concentrations at
The assessment of interleukin (IL)-8 in EBC has received less attention. Izquierdo et al. 4 described lower concentrations of IL-8 in EBC from emphysematous subjects (0.34 pg·mL–1), compared with chronic bronchitics (2.32 pg·mL–1) and controls (3.32 pg·mL–1). Zihlif et al. 6 suggested that IL-8 concentrations were undetectable in both children with primary ciliary dyskinesia and controls. Simpson et al. 7 described detectable IL-8 in smokers but questioned the validity of the ELISA, whereas Sack et al. 1 were able to measure IL-8 in EBC using a multiplex system. However, very little information is available concerning other proteins such as The present study sought to determine whether it was possible to validate a series of assays for use in the measurement of mediators in EBC. The assays under investigation were: commercial ELISAs for LTB4 (GE Healthcare Life Sciences, Little Chalfont, UK and Cayman Chemical Company) and for IL-8 and SLPI (R&D Systems Europe Ltd, Abingdon, UK); an in-house ELISA for AAT; and a chromogenic activity assay for MPO. In addition, results in EBC were compared with those found in matched, spontaneous, sol phase sputum samples from the same patient on the same day, to determine whether the EBC level reflected the results of airway secretions.
Study subjects In total, 61 subjects were recruited for the study: 12 patients with bronchiectasis (confirmed on high-resolution computed tomography), 19 patients with COPD, and 30 normal subjects with normal spirometry and no history of lung disease. All subjects were assessed in a stable state (no antibiotics or oral corticosteroids for 2 months). Of the 19 COPD patients, 12 fulfilled standard criteria for the diagnosis of chronic bronchitis (CB; daily sputum production for 3 months of 2 consecutive yrs) 8. Patients with underlying immune deficiency, allergic bronchopulmonary aspergillosis or cystic fibrosis were excluded from the study.
Assessment and investigations Following EBC collection, all subjects underwent spirometry with reversibility (20 min before and after inhalation of 400 µg of salbutamol) 10. Post-bronchodilator forced expiratory volume in one second (FEV1) was expressed as percentage of the predicted normal reference values 11.
Assays on sputum sol phase and EBC
The LLD for an assay was defined as 2 SD above the mean signal for 20 sample blanks. The LLQ was defined as the point at which both intra- and interassay CV and spike recovery became acceptable for each assay (<12%). This concept is demonstrated for the SLPI ELISA in figure 1.
The methods for LTB4 (GE Bioscience), SLPI and MPO have been described in detail previously 12. The second commercial LTB4 ELISA (Cayman Chemical Company) was used according to the manufacturers instructions. IL-8 was measured by ELISA using a commercially available kit (R&D Systems Europe Ltd). The sol phase intra-assay CV was <8.5% throughout the working range of the assay. Sputum samples spiked with IL-8 resulted in >88% recovery. AAT was measured by an in-house ELISA relative to a commercially available serum standard (The Binding Site Limited, Birmingham, UK). In brief, 200 µL of anti-human AAT (The Binding Site Limited) in 0.05 M sodium carbonate/bicarbonate pH 9.6 was added to a Nunc MAXISORP (Loughborough, UK) microtitre plate and incubated overnight at 4°C. The plate was then washed three times with PBS containing 1% (v/v) Tween 20 (Sigma-Aldrich Company Ltd, Poole, Dorset, UK) and 0.5% (w/v) dried skimmed milk (PBS-T), 200 µL of standard or sample was added to the plate, and it was then incubated for 1 h at 37°C. The plate was again washed three times with PBS-T, then 200 µL of anti-human AAT peroxidase conjugate (The Binding Site Limited) in PBS-T was added to each well, and the plate was incubated for 1 h at 37°C. Following three more washes with PBS-T, 200 µL of 3,3',5,5'-tetramethylbenzidine ELISA substrate solution (Sigma-Aldrich Company Ltd) was then added to each well and incubated for 10 min at 25°C. The reaction was stopped with 50 µL of 0.1 M H2SO4 and the plate was read at 450 nm with a 570 nm correction, and the AAT concentration obtained by interpolation from the standard curve. The sol phase intra-assay CV was <10.2% throughout the working range of the assay and sputum samples spiked with AAT gave >85% recovery.
Statistics
Baseline characteristics Baseline characteristics are shown in table 1
Assays on sputum sol phase and EBC All assays had an intra- and interassay CV of <15% in the linear part of the standard curve and >85% spike recovery.
LTB4
The present authors also assessed a second LTB4 assay (Cayman Chemical Company) and found that the LLD for this assay was 9.2 pg·mL–1, which was consistent with the value reported by the manufacturer (
IL-8
SLPI
AAT
MPO
Correlation with sputum
The collection of EBC is thought to provide a valuable noninvasive technique to measure inflammatory mediators within the airways. However, the assays described in the present study show that extreme care needs to be taken when interpreting mediator results in this fluid. All of the ELISAs described confirm poor reproducibility for values derived from below the linear part of the sigmoid curve of standard quantity to signal described by an ELISA. This is understandable, as small variations in the optical density at this point can result in large changes in the value derived by interpolation and confirmed by variable and unreliable spike recovery results. The most accepted definition of LLQ is the lowest concentration that can be measured with a definite level 14. The LLQ described in the present study was the value above which the intra-assay CV was shown to be <12% and the spiked mediator recovery was >80%, and assay characteristics were clearly linear. Samples from only 19.6% of the subjects in the present study were within these criteria, and for only one assay (AAT). There has been very little data published regarding the validation of assays used in EBC at very low concentrations. SLPI, AAT and MPO activity measurements have not previously been reported in EBC. In the present study, the sol phase of spontaneous sputum was chosen as a read-out for mediator concentrations found within the lungs. Currently, no obvious "gold standard" exists for reference of the EBC mediator concentrations with those found within the lungs 15. EBC reflects air from the whole bronchial tree. However, airway secretions (in particular spontaneous sputum) contain high levels of mediators. As EBC and sputum were collected from the same patients, this was felt to be the best comparison. In addition, the data by Biernacki et al. 16 demonstrated that EBC LTB4 is increased in exacerbation of COPD and falls upon resolution. Since exacerbations of COPD are events affecting the bronchial tree, and since similar changes occur in sputum, this is the most appropriate sample for comparative purposes 17. It is possible that EBC also reflects changes in the distal airways and hence comparison with bronchoalveolar lavage (BAL) would be appropriate, even in light of the lack of relationship with BAL demonstrated by Jackson et al. 15. However, it seems inappropriate to assess this until valid assays are identified for EBC. In order to enable reliable mediator determination, it might be argued that concentrating the EBC is necessary, although such procedures may also prove inaccurate due to protein loss 18 and other confounding factors, such as unknown stability of LTB4 in EBC when lyophilised 9. The results described in the current study suggest that EBC cannot be used to determine mediator concentrations either within a study or as a diagnostic tool, via the present assays and collection techniques. Although the current results question the validity and conclusions drawn from other studies involving the mediators assessed here, EBC may still prove a useful tool. Assays, such as IL-6, might detect quantifiable levels in EBC, and other biomarkers may prove measurable by current methodologies. The use of other collection equipment, or the application of coatings to the collection equipment to prevent binding of mediators, may be necessary. For example, Rosias et al. 19 have shown that glass and silicone were superior to aluminium, polypropylene and Teflon when measuring 8-isoprostane and albumin levels in EBC. Liu et al. 20 demonstrated that the total protein levels in EBC were higher when using Ecoscreen® when compared with glass and RTubeTM in collection devices, and that Ecoscreen® might prevent mediator binding. Leung et al. 5 have demonstrated that cysteinyl LTs and LTB4 were poorly correlated when collected using Ecoscreen® and RTubeTM. However, the effect of differing collection devices upon specific protein inflammatory mediators is unknown. Alternatively, it might be possible to prevent absorption of mediators upon the collection equipment by pre-blocking with agents such as bovine serum albumin, although this might in itself cause reproducibility issues. The presence of proteinases, receptors, antibodies and other interfering proteins, along with variations in pH and protein concentration, ensure that sputum is a challenging matrix for mediator measurement. However, the high mediator concentrations present often enable deleterious effects upon ELISAs to be substantially negated by dilution, as has been shown for secretory proteinase inhibitor in the presence of neutrophil elastase 21. Critically, this is not possible in EBC at present. The alternatives are to develop more sensitive ELISA assays, proteomics or mass spectroscopy measurements. However, with more sensitive assays a balance in the signal/noise ratio will need to be taken into account 9. The results of the present study indicate that the performance of any assay must be fully characterised in order to interpret the results. The current study investigated both ELISA and chromogenic substrate techniques; however, the validation methodologies are also critical for all methodologies used for the assessment of mediators within exhaled breath condensate.
A statement of interest for R.A. Stockley can be found at www.erj.ersjournals.com/misc/statements.shtml