Copyright ©ERS Journals Ltd 2008 Biomarker reproducibility in exhaled breath condensate collected with different condensersDepts of 1 Paediatric Pulmonology, Caphri Research Institute, 7 Paediatrics and 5 Clinical Chemistry and Clinical Proteomics, University Hospital Maastricht and Depts of 3 Methodology and Statistics and 8 General Practice, Caphri Research Institute, 4 Pharmacology and Toxicology, Maastricht University, Maastricht, and 2 Dept of Paediatrics, Maasland Hospital, Sittard, and 6 Dept of Immunology, University Medical Centre Wilhelmina Children's Hospital, Utrecht, The Netherlands. CORRESPONDENCE: P. P. Rosias, Dept of Paediatric Pulmonology, Caphri Research Institute, University Hospital Maastricht, PO Box 5800, 6202 AZ Maastricht, The Netherlands. Fax: 31 433875246. E-mail: p.rosias{at}orbisconcern.nl Keywords: Condenser, cytokines, exhaled breath condensates, hydrogen peroxide, 8-isoprostane, multiplex array
Received: June 19, 2007
Optimal collection and analysis of exhaled breath condensate (EBC) are prerequisites for standardisation and reproducibility of assessments. The present study aimed to assess reproducibility of EBC volume, hydrogen peroxide (H2O2), 8-isoprostane and cytokine measurements using different condensers, including a newly developed glass condenser. At four points in time, 30 healthy subjects performed sequential EBC collections randomly using the following four condensers: glass, silicone, EcoScreen® (Erich Jaeger GmbH, Hoechberg, Germany) and an optimised glass condenser. In small EBC samples, H2O2 was measured by spectrophotometer, 8-isoprostane by enzyme immunoassay, and cytokines by multiplexed xMAP® technology (Luminex Corporation, Austin, TX, USA).
The optimised glass condenser yielded significantly more EBC volume (median 2,025 µL, interquartile range 1,600–2,525). The reproducibility of EBC volume, yielded by the new glass condenser, was comparable with EcoScreen® (19–20 coefficients of variation (CV)%), but was significantly better compared with silicone and glass (29–37 CV%). The new condenser was associated with significantly more detections of H2O2, 8-isoprostane, interleukin-2, -4, -5 and -13, and tumour necrosis factor- In conclusion, significantly more exhaled breath condensate volume and biomarker detections were found using the optimised glass condenser, including higher 8-isoprostane levels. However, biomarker reproducibility in exhaled breath condensate in healthy adults was not influenced by the type of condenser. The collection of exhaled breath condensate (EBC) is a noninvasive, safe technique with which to obtain direct samples from the lower respiratory tract, without disturbing an ongoing inflammation 1–3. Analysis of EBC reveals the presence of inflammatory markers, such as eicosanoids, hydrogen peroxide (H2O2) and cytokines 1–3. Although the American Thoracic Society and European Respiratory Society Task Force on EBC published general methodological recommendations on the collection and analysis of EBC, there are still some unresolved methodological pitfalls, as illustrated by the use of various nonstandardised collection systems 1–3. Optimal condensate collection, and optimal biomarker detection and measurement in EBC are reciprocal prerequisites for any standardisation. However, current condensation systems are suboptimal, with relatively short-measured, open-ended designs and loss of noncondensed exhaled breath, as reflected by variable EBC volumes and biomarker reproducibility 1. Logically, modification by using guided breath flows, enlarged condensation surface, and optimised condensate recovery may improve condensation. Moreover, current designs have different inner coatings, featuring different adhesive interactions with exhaled markers 4. Recently, loss of biomarker within the sampling system was demonstrated, both in vitro and in vivo, for 8-isoprostane and albumin, at the expense of nonglass condenser systems, and in favour of glass and silicone condensers 4. To assess clinical relevance, a study on the reproducibility of these biomarker measurements in vivo, using different condenser coatings including glass, is needed 4. Furthermore, conventional biomarker assays are not always suitable for use in even large sample volumes of condensate, a biofluid highly "diluted" by water vapour 5, whereas new analytical techniques are rapidly emerging and may offer new perspectives 6, 7. Recently, multiplexed cytometric bead array was used in children to simultaneously measure different cytokines in only 50-µL condensate samples; however, the detection level did not reach 50% 6. Liquid bead-based multiplexing xMAP® technology (Luminex Corporation, Austin, TX, USA), based on flow cytometry and (faster) liquid microspheres reaction kinetics, is less laborious, highly sensitive and specific, allows simultaneous measurements in small sample sizes, and improves interarray reproducibility 7. Therefore, it was hypothesised that optimised condensate collection with minimal adhesive properties improves the reproducibility of different measurements in EBC.
The aim of the present study was to assess the effect of four different condensers (glass, silicone, EcoScreen® (Erich Jaeger GmbH, Hoechberg, Germany) and a new, optimised glass condenser) on the reproducibility of breath condensate volume, and the detection, concentration and reproducibility of inflammatory biomarkers, including H2O2, 8-isoprostane, interleukin (IL)-2, -4, -5, -6, -8, -10 and -13, and tumour necrosis factor (TNF)-
Study subjects A total of 30 eligible healthy nonasthmatic adult volunteers were recruited (table 1 15 min. Exclusion criteria were as follows: history of asthma, upper or lower airway infection, and use of antibiotics, corticosteroids, cromoglycate, nedocromil, theophylline or leukotriene antagonist. Nonasthmatic healthy adults were chosen to eliminate possible confounding factors attributable to heterogeneous disease expression and/or variability of disease control.
Study design To assess within-day, between-day and between-week reproducibility, each volunteer was asked to perform the following four tests. Test 1: on the first day in the morning; test 2: on the first day in the afternoon; test 3: the second day at the same time as test 1; and test 4: 1 week after test 1 at the same time. In turn, each test consisted of four sequential EBC collections, using random different types of condenser.
EBC collection
To perform one EBC collection, each subject was asked to exhale tidally, while using a nose-clip, through a mouthpiece and two-way nonrebreathing valve connected with the condenser, during a fixed period of 15 min.
Condensate sample processing H2O2 was measured in 50-µL EBC, in duplicate by spectrophotometer (UV-VIS Lambda 10 Spectrometer; Perkin Elmer, Shelton/Norwalk, CT, USA) with a lower detection limit of 0.05 µM, as described previously 8. Isoprostane was measured in 100 µL of EBC, by specific enzyme immunoassay (Cayman Chemical, Ann Arbor, MI, USA), which was modified to reach a lower limit of detection of 1.0 pg·mL–1. The standard curve of this assay ranged 250–1.95 pg·mL–1. It was possible to report 8-isoprostane values as low as 1.0 pg·mL–1, as a logit/log transformation was used. Isoprostane recovery experiments were performed, and coefficients of variation (CV) of the absorption signals were assessed. When spiking for the lower 8-isoprostane values of 3.9 and 7.8 pg·mL–1, an isoprostane recovery of 92% (CV of concentration 16%) and 95% (CV of concentration 15%) were found, respectively. A CV (of absorption signals) <15% was considered highly acceptable. Hence, the corresponding CV of concentrations may be higher. Therefore, the CVs found in the 8-isoprostane recovery experiments using these low values were considered good. Standard curves, patient samples and quality control samples of 2.5 and 10 pg·mL–1 8-isoprostane were assayed in triplicate. Intra-assay variation of standard curves, patient samples and quality control samples had to be <15%, otherwise all samples measured in that assay were excluded and reanalysed. Finally, in all accepted samples, 8-isoprostane concentrations were determined from mean absorption signal intensities.
Cytokines were measured in 50-µL EBC, using the liquid bead-based multiplexing xMAP® technology. Multiplex immunoassay was performed as described previously 7, 9, 10. The corresponding lower limits of cytokine and chemokine detection were as follows (in pg·mL–1): IL-2 (1.0), IL-4 (1.2), IL-5 (1.2), IL-6 (0.4), IL-8 (1.1), IL-10 (1.2), IL-13 (1.0) and TNF-
Statistics CV = (SD/mean) x 100% (1) The CV was calculated as the mean of individual CVs, calculated over the two or four relevant measurements. Within-subject CVs were used as data distribution was normal. When the distribution of individual CVs was not normal, the nonparametric Friedman test was used to see if these CVs were different between coatings. ANOVA was used to test for differences among normally distributed repeated measures. The Chi-squared test was used to statistically evaluate the differences between proportions for four groups in a data set. Condensate samples with a biomarker concentration below the lower detection limit, i.e. in strictu sensu negative detections, were not considered as missing values because they actually informed that marker concentrations were below the lower detection limit. Therefore, these negative detections were given an arbitrary value between zero and the lower detection limit as follows: 0.025 µM for H2O2, 0.1 pg·mL–1 for 8-isoprostane and 0.1 pg·mL–1 for cytokines. Samples were defined as missing if the EBC volume, yielded after 15 min collection time, was either zero or insufficient to analyse.
Power calculation
Ethics
EBC volume In five out of 480 manoeuvres performed, the present authors were unable to collect condensate due to erroneously connected tubing. EBC collection by the optimised glass condenser yielded significantly more median condensate volume compared with the other condensers (p = 0.001, Friedman test; table 2
H2O2 measurements in EBC Overall, 29% of H2O2 measurements were missing (table 3
8-Isoprostane measurements in EBC Overall, 13% of 8-isoprostane measurements were missing (table 4
Cytokine measurements in EBC Overall, 20% of cytokine measurements were missing (table 5 (64%). The optimised glass condenser had significantly more positive detections of IL-2, -4, -5, -13 and TNF- , compared with silicone and glass (p<0.050, Chi-squared test), and more IL-5 and -13 detections, compared with EcoScreen® (p<0.021, Chi-squared test). Cytokine concentrations were not significantly different between the four condenser types (p>0.113, Friedman test or ANOVA, respectively; table 6 (8–22%). When CVs were evaluated using the positive values only (without negative and/or arbitrary values), again no differences between condensers were found (data not shown).
The present authors have demonstrated that EBC volume and the detection of biomarkers were significantly influenced by the condenser system, in favour of the new glass condenser, whereas biomarker reproducibility was not influenced by the type of condenser. The proposed optimised glass condenser yielded significantly more condensate volume compared with the silicone, glass and EcoScreen® condensers. Reproducibility of EBC volume was comparable for the new condenser and EcoScreen®, and was significantly better compared with the other two condenser types.
In EBC collected with this new condenser, significantly more positive H2O2, 8-isoprostane, IL-2, -4, -5 and -13, and TNF- Variations in biomarker measurements in EBC may be attributed to variations in the dilution and/or quality of condensate that is influenced by different collection techniques and procedures, sample processing and storage conditions, and/or sensitivity of the analytical techniques used 17. The levels of the highly volatile H2O2 in EBC may be susceptible to different cooling temperatures during collection 18, circadian rhythm 19, flow dependency 20, different methods of measurement with widely varying values (even in healthy subjects) close to lower detection limits 3, 17, 21, 22, and high chemical reactivity with salivary and exhaled compounds by which (some) H2O2 is consumed over time during collection and storage 3, 17, 22. Isoprostanes are relatively stable end-products of in vivo arachidonate peroxidation and are measured in EBC by immunoassay, which might be influenced by cross-reactivity with closely related substances 17. Other confounding factors may be age, diets rich in antioxidants and smoking habits 17. The use of CV is not always the ideal way to express variability: when mean values are low, CV values can be abnormally high. Therefore, results in the present study were also expressed as SD values and intraclass correlation coefficients. When using both of these alternative expression methods, results were comparable: no difference in biomarker variability between condensers was found (data not shown). Furthermore, biomarker levels were assessed in condensate originating from healthy adults and, thus, may have been more pronounced in a steroid-naïve population with documented chronic respiratory inflammation and comparable levels of disease control. Hence, increased mean concentrations and lowered CVs in diseased subjects could be expected. Conversely, with these very low concentrations of cytokines and 8-isoprostane, an influence of analytical variability cannot be ruled out. Cytokines were simultaneously measured by xMAP® technology in small, 50-µL EBC samples. The overall level of detection was 46–97%, which was much better compared with cytometric bead array in small samples in children (<50%), and in large 1,000–2,000-µL lyophilisated samples in adults (3–100%) 6, 23–26. When compared with conventional (solid-phase) immunoassays, multiplexed immunoassays detect bioactive and inactive molecules, have a growing analytical range, are rapid (take hours instead of days to perform), have good precision (CV 10–15%), are not interfered with by drugs, and have simple protocols 7. Currently, different nonstandardised techniques to collect EBC are in use 1–3. All systems are based on the cooling of exhaled breath, whereas their design may vary from immersed plastic tubing over glass distilling columns, to commercial systems, such as the lamellar Teflon-like EcoScreen® condenser, the hand-held disposable polypropylene RTubeTM (Respiratory Inc., Charlottesville, VA, USA), and the thermostatically controlled polyethylene Turbo-Deccs (ItalChill, Parma, Italy) 1–3, 27. These designs implicate relatively short-measured and open-ended systems that tolerate the needless loss of noncondensed exhaled breath, whether initially, or after a prolonged collection time. Moreover, biomarkers may also be lost within these collection systems, as recently demonstrated for 8-isoprostane and albumin measurements, both in vitro and in vivo, in nonglass condensers (including EcoScreen®) 4. This superiority of glass coatings may be mainly related to the behaviour of water as bipolar vehicle 4. Three other studies report the influence of sampling systems on biomarkers in EBC 28–30. Tufvesson and Bjermer 28 proposed to coat EcoScreen® collection surfaces with bovine serum albumin and Tween-20, to measure cytokines and eicosanoids, respectively. However, they also reported possible false-positive (eicosanoid) results, and the need of sample concentration, by vacuum centrifugation, prior to analysis. Soyer et al. 29 found significantly higher cysteinyl leukotrienes and eotaxin using EcoScreen®, compared with RTubeTM, due to susceptibility of pre-cooled RTubeTM sleeves to (increased) ambient temperatures during collection, and due to different materials that could affect sample recovery. Prieto et al. 30 compared RTubeTM and EcoScreen®, and reported that EBC pH values are dependent on the collection device used. Furthermore, the EcoScreen® has been associated with deposition of frozen condensate on its lamellar walls. For these reasons, an optimised glass condenser system was developed by the present authors. The new glass condenser had an improved condensation process and condensate recovery, using an inclined and enlarged condensation surface, with a condensate sweeping plunger, having tangentially and axially guiding breath-flow channels. In the present study, these improvements resulted in significantly increased EBC volumes, and increased biomarker detections with the new glass condenser compared with silicone, glass and EcoScreen®. This suggests both an improved condensation process and an increased opportunity to perform a broad spectrum of analyses. Moreover, the optimised glass and EcoScreen® condensers were both significantly associated with less variation in the generated EBC volume compared with the other condensers, thereby reducing a possible confounding influence of the variable quantity of EBC collected over a given time, and even within individuals. Optimisation of EBC collection using the modified new glass condenser with statistically equivalent CVs compared with the commercial EcoScreen®, and optimisation of EBC analysis using rapid, multiplexed measurement of cytokines in small EBC sample sizes may, in combination, open a window of opportunities, even in strained collection procedures, such as in young or dyspnoeic subjects (with less-sustained efforts to cope with sampling procedures), by allowing the search for, and identification of, particular profiles of different exhaled markers involved in the regulation of chronic respiratory inflammation for diagnostic and monitoring purposes.
In conclusion, the optimised glass condenser yielded significantly more exhaled breath condensate volume, and with good reproducibility. Furthermore, significantly more positive detections of hydrogen peroxide, 8-isoprostane, interleukin-2, -4, -5 and -13, and tumour necrosis factor-
None declared.
The authors wish to thank J. Suykerbuyk (Dept of Paediatrics, University Hospital Maastricht, Maastricht, the Newtherlands) for her technical support in the collection of exhaled breath condensate; M. Meers (Dept of Clinical Chemistry, University Hospital Maastricht) for her technical support in the measurement of 8-isoprostane; N. van Uden (Dept of Immunology, University Medical Centre Wilhelmina Children's Hospital, Utrecht, the Netherlands) for her technical support in the measurement of cytokines; and, last but not least, the volunteers for their participation.