Eur Respir J 1998; 11: 1121-1127
Copyright © ERS Journals Ltd 1998
Influence of correction of flow limitation on continuous positive airway pressure efficiency in sleep apnoea/hypopnoea syndrome
JC Meurice,
J Paquereau,
A Denjean,
F Patte,
and
F Series
We evaluated the efficiency of two different treatment procedures with continuous positive airway pressure (CPAP) on sleep, nocturnal breathing characteristics and daytime vigilance in 18 newly diagnosed patients with untreated sleep apnoea/hypopnoea syndrome (SAHS) randomly allocated to two different groups. In group I, the positive pressure (PP) level was set to suppress flow limitation (PFL), while in group II the PP was set at a level that eliminated only apnoea/hypopnoea and snoring (PAHS). At the end of a 3 week period of home CPAP therapy, a follow-up sleep study, vigilance and cognitive tests were made. Overall, PFL was significantly higher than PAHS values (PFL: 10.42.6 cmH2O; PAHS: 8.9+/-2.6 cmH2O; p<0.01, mean+/-SD). We found no difference in sleep quality, nocturnal saturation and apnoea/hypopnoea index, or in daytime vigilance tests between the two groups at the end of the treatment period. However, there was a significantly greater scattering in the changes of sleep latency in group II than in group I. This was associated with a significant difference in the daily duration of nasal CPAP use between the two groups (group I: 7.29+/-0.95 h x day(-1); group II: 6.01+/-0.94 h x day(-1); p=0.01) and with a positive correlation between final maintenance of wakefulness test values and the duration of CPAP use (p<0.05; r=0.55). These results tend to show that correcting flow limitation is associated with a higher observance and a more important efficiency in normalizing daytime vigilance than with conventional nasal continuous positive airway pressure.
This article has been cited by other articles:

|
 |

|
 |
 
M. H. Sanders, J. M. Montserrat, R. Farre, and R. J. Givelber
Positive Pressure Therapy: A Perspective on Evidence-based Outcomes and Methods of Application
Proceedings of the ATS,
February 15, 2008;
5(2):
161 - 172.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
D. Pevernagie, J. F. Masa, J. C. Meurice, R. Farre, O. Marrone, and D. Rodenstein
Treatment of obstructive sleep-disordered breathing with positive airway pressure systems
Eur. Respir. Rev.,
December 1, 2007;
16(106):
125 - 131.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
R. K. Kakkar and R. B. Berry
Positive Airway Pressure Treatment for Obstructive Sleep Apnea
Chest,
September 1, 2007;
132(3):
1057 - 1072.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
F. Lofaso, G. Desmarais, K. Leroux, V. Zalc, R. Fodil, D. Isabey, and B. Louis
Bench evaluation of flow limitation detection by automated continuous positive airway pressure devices.
Chest,
August 1, 2006;
130(2):
343 - 349.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
A. Noseda, C. Kempenaers, M. Kerkhofs, S. Braun, P. Linkowski, and E. Jann
Constant vs Auto-Continuous Positive Airway Pressure in Patients With Sleep Apnea Hypopnea Syndrome and a High Variability in Pressure Requirement
Chest,
July 1, 2004;
126(1):
31 - 37.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. T. Kuna
Can Continuous Positive Airway Pressure Be Self-Titrated?
Am. J. Respir. Crit. Care Med.,
March 1, 2003;
167(5):
674 - 675.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
E Verin, T Similowski, and F Series
Effects of continuous positive airway pressure on upper airway inspiratory dynamics in awake patients with sleep-disordered breathing
J. Physiol.,
January 1, 2003;
546(1):
279 - 287.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
W. J. RANDERATH, O. SCHRAEDER, W. GALETKE, F. FELDMEYER, and K.-H. RUHLE
Autoadjusting CPAP Therapy Based on Impedance Efficacy, Compliance and Acceptance
Am. J. Respir. Crit. Care Med.,
March 1, 2001;
163(3):
652 - 657.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M.-P. d'Ortho, V. Grillier-Lanoir, P. Levy, F. Goldenberg, E. Corriger, A. Harf, and F. Lofaso
Constant vs Automatic Continuous Positive Airway Pressure Therapy : Home Evaluation
Chest,
October 1, 2000;
118(4):
1010 - 1017.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
J. R. BADIA, R. JOHN KIMOFF, E. BALLESTER, L. HERNÁNDEZ, M. ROTGER, D. NAVAJAS, and J. M. MONTSERRAT
Clinical Application of the Forced Oscillation Technique for CPAP Titration in the Sleep Apnea/Hypopnea Syndrome
Am. J. Respir. Crit. Care Med.,
November 1, 1999;
160(5):
1550 - 1554.
[Abstract]
[Full Text]
|
 |
|
Copyright © 1998 by the European Respiratory Society.
|