ERJ
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Kampelmacher, M.
Right arrow Articles by Winckers, E.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Kampelmacher, M.
Right arrow Articles by Winckers, E.
Eur Respir J 1997; 10: 1341-1344
Copyright © ERS Journals Ltd 1997


Original Articles

Instrumental variability of respiratory blood gases among different blood gas analysers in different laboratories

MJ Kampelmacher, RG van Kesteren, and EK Winckers

The aim of this study was to test the hypothesis that differences in oxygen tension (PO2) and carbon dioxide tension (PCO2) values from measurements performed on different blood gas analysers in different laboratories are clinically insignificant. Samples of fresh whole human tonometered blood (PO2 8.1 kPa (60.8 mmHg); PCO2 5.3 kPa (39.9 mmHg)) were placed in airtight glass syringes and transported in ice-water slush. Blood gas analysis was performed within 3.5 h by 17 analysers (10 different models) in 10 hospitals on one day. The mean of the differences between the measured and target values was -0.01+/-0.19 and 0.21+/-0.13 kPa (-0.06+/-1.45 and 1.55+/-1.01 mmHg) for PO2 and PCO2, respectively. The mean of the differences between two samples on one analyser was 0.06+/-0.06 and 0.04+/-0.03 kPa (0.47+/-0.48 and 0.29+/-0.24 mmHg), respectively. For PO2 and PCO2 the interinstrument standard deviations (s(b)) were 0.18 and 0.13 kPa (1.38 and 0.99 mmHg), respectively, whereas the intra-instrument standard deviations (s) were 0.06 and 0.03 kPa (0.47 and 0.26 mmHg), respectively. Both for PO2 and PCO2 the ratios of s(b)2 and s2 were statistically significant (analysis of variance (ANOVA) p<0.001). The standard deviations of a random measurement on a random analyser were 0.19 and 0.14 kPa (1.46 and 1.02 mmHg) for PO2 and PCO2, respectively. We conclude that the variability in measurement of blood gas values among different blood gas analysers, although negligible, depends much more on inter- than intra-instrument variation, both for oxygen tension and carbon dioxide tension. Technical improvements and adequate quality control programmes, including tonometry, may explain why the variability in blood gas values depends mainly on errors in the pre-analytical phase.


This article has been cited by other articles:


Home page
Arch. Dis. Child. Fetal Neonatal Ed.Home page
C Morgan, S J Newell, D A Ducker, J Hodgkinson, D K White, C J Morley, and J M Church
Continuous neonatal blood gas monitoring using a multiparameter intra-arterial sensor
Arch. Dis. Child. Fetal Neonatal Ed., March 1, 1999; 80(2): 93F - 98.
[Abstract] [Full Text]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1997 by the European Respiratory Society.